The authors report explicitly state that the system's resolution does not match clinical CT or MRI. The elevational resolution is 15–25mm, meaning each slice is effectively a thick 2D section rather than the fine isotropic 3D volume MRI provides. MRI also delivers far richer soft-tissue contrast; this device produces only three contrast types (reflectivity, speed of sound, attenuation), and because it uses a low 1MHz frequency, the reflection images come primarily from tissue boundaries rather than from internal tissue texture.

I could see this being valuable for adipose tissue mapping or fatty-liver monitoring at a large scale, as the machines would be significantly cheaper, but this isn't some revolutionary magic bullet like the Twitter post is insinuating.