>my HP 3458A can do a small handful of 8.5 digit DCV measurements per second.

I thought they had like 100k samples a second. Is it really sufficient for 8.5 precision?

Edit: perhaps measurement should be confined only to relative finish between the participants, e.g. finish within 1ms should be considered the same (regardless if they fall in the same hundreds of a second bucket)

2 comments

I don’t keep mine in cal (or even powered up all the time), so I can’t claim full volt-nut status here, but my understanding is the 100K/sec 4.5 digit measurement rate is limited by communications not by sampling rate. (The 5.5 digit rate is 50K/sec, implying the 4.5 digit rate limit is not in the measurement but more likely in the GPIB.)

On thing I know with perfect precision: HP/Keysight knows more than me about this. :)

interestingly enough, for stage races, that is how things work to some degree for the "pack" finish, just not the guys going for the day's podium